Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-35
4
Operation
SatShare Operation (continued)
CAUTIONS:
SATSHARE SIGNALS ARE IDEAL SIMULATED WAVEFORMS CORRESPONDING
TO THE CALCULATED SATURATION AND PULSE RATE VALUES AND DO NOT
CONTAIN ALL OF THE INFORMATION CONTAINED IN PHYSIOLOGICAL WAVEFORMS.
THE MULTIPARAMETER PATIENT MONITOR DECODES THESE SIGNALS INTO
SATURATION AND PULSE RATE VALUES.
DURING SATSHARE OPERATION, THE AUDIBLE ALARMS MAY BE MUTED ON THE
RADICAL-7 PULSE CO-OXIMETER. WHEN THE AUDIBLE ALARM IS MUTED (INDICATED
BY A BELL WITH A SLASH THROUGH IT) ON THE RADICAL-7 PULSE CO-OXIMETER,
USE THE MULTIPARAMETER MONITOR FOR AUDIBLE ALARM INDICATION.
THE (RADICAL-7) SpO
2
AND PULSE RATE AUDIBLE ALARMS MAY BE DISABLED
WHEN THE RADICAL-7 IS CONFIGURED WITH RAINBOW PARAMETERS/
MEASUREMENTS. REFER TO SECTION 4, ALARMS.
THE RADICAL-7 REVERTS TO THE INTERFACE ALARM SpO
2
/BPM “Yes” SETTING DURING
POWER INTERRUPTIONS OR WHEN THE SATSHARE CONNECTION IS LOST OR THE
Instrument BECOMES SEPARATED FROM THE DOCKING STATION. THIS ENSURES THAT
THE RADICAL-7 PROVIDES AUDIBLE ALARMS FOR SpO
2
AND PULSE RATE WHEN THE
CONNECTION TO THE INTERFACED SYSTEM BECOMES COMPROMISED.
DURING SATSHARE OPERATION DO NOT USE THE PLETHYSMOGRAPHIC
WAVEFORM DISPLAY ON THE MULTIPARAMETER MONITOR FOR DIAGNOSTIC
PURPOSES. INSTEAD, USE THE PLETHYSMOGRAPHIC WAVEFORM DISPLAYED ON
THE RADICAL-7 SCREEN.
TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANY EQUIPMENT
TO THE SATSHARE CONNECTOR UNLESS THE RADICAL-7 IS CONNECTED TO THE
AC MAINS POWER SUPPLY.
ONLY USE A SATSHARE CABLE THAT HAS A FERRITE BEAD INSTALLED.
ONLY SPO
2
AND PULSE RATE CAN BE DISPLAYED ON THE MULTIPARAMETER
MONITOR WITH SATSHARE.
To return from SatShare operation to normal standalone operation, simply disconnect the
SatShare cable from the patient monitor or disconnect the SatShare cable from the SatShare
connector on the back of the Radical-7 Pulse CO-Oximeter.