Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 9-5
9
Service / Maintenance
Performance Verification
To test the performance of the Radical-7 Pulse CO-Oximeter following repairs or during routine
maintenance, follow the procedure outlined in this section. If the Radical-7 Pulse CO-Oximeter
fails any of the described tests, discontinue its use and correct the problem before returning the
instrument back to the user.
Before performing the following tests place the Radical-7 Pulse CO-Oximeter Handheld into the
Docking Station, connect the Radical-7 to AC power and fully charge the Radical-7 Pulse CO-Oximeter
Handheld battery. Also disconnect any patient cables or pulse oximetry probes, as well as SatShare,
serial or analog output cables from the instrument. Set the Radical-7 Pulse CO-Oximeter to normal
operating mode by selecting the Home Use parameter in the General Menu to “No”.
Power-On Self-Test:
1. Connect the monitor to AC power and verify that the AC Power Indicator is lit.
2. Turn the monitor on by depressing the Power/Standby Button. Within 5 seconds all
available LEDs are illuminated, a 1-second beep tone sounds, and the Masimo SET logo
is displayed.
3. The blue Docking Indicator LED is illuminated and the Radical-7 begins normal operation.
Key Press Button Test:
With the exception of the Power/Standby Button, press each soft key button and verify that the
Radical-7 acknowledges each key-press with an audible beep tone or by indicating a change
on the display.
Alarm Limit Test:
1. With the monitor turned on, select the Menu Access key and enter the Alarm menu.
Change the High SpO
2
Alarm parameter to a value two points below the currently selected
value, and accept the change.
2. Verify that the newly set parameter is shown on the Saturation Alarm Limit Display, next to
the SpO
2
or pulse rate measurement display.
3. Return the High Saturation Alarm parameter to its original setting.
4. Repeat steps 1 to 3 for the following alarm parameters:
■ Low SpO
2
■ High and Low Pulse Rate
■ High SpMet
■ High SpCO
■ High and Low SpHb
5. Reset the alarm limits again to the original settings.
Display Contrast Test:
1. With the monitor turned on, select the Menu Access key and enter the Display menu.
Change the Contrast parameter by scrolling through the contrast settings.
2.
Return the Contrast setting to the original value, or a value that allows maximum viewing contrast.
3. Exit the Menu system and press and hold down the Backlight/Contrast button for several
seconds. The display will scroll again through all the contrast settings.
4.
Release the Backlight/Contrast button again when the display shows maximum viewing contrast.