Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-25
4
Operation
Service
NOTE: The Service menu selections are only available when the Radical-7 Pulse CO-Oximeter
Handheld is interfaced to the Docking Station.
Only qualified Biomedical or Clinical Engineering department personnel should access the
service menu. See Section 4, Password Operation, on how to enter the password.
MENU ITEMS DESCRIPTION
HANDHELD
BATTERY
DISCHARGE
To deep discharge the Handheld battery, select this menu item. See
Section 9, Battery Operation and Maintenance, for more information.
DS BATTERY
DISCHARGE
To deep discharge the optional Docking Station battery, select this
menu item. See Section 9, Battery Operation and Maintenance, for
more information.
The discharge cycle will take approximately 16 hours to complete for the Handheld battery. The
Docking Station battery will take approximately 30 hours to complete. A message will appear in
the service screen when the discharge cycle is complete. The batteries will be fully charged after
completion of the cycle.
When deep discharge is started, the backlight will automatically revert the default handheld battery
powered level. Wait until the message changes from "In Progress" to "Done".
NOTE: In order for the discharge cycle to be properly completed, AC power must be supplied to
the instrument throughout the cycle.
WARNING:
WHEN DEEP-DISCHARGING THE HANDHELD OR DOCKING STATION
BATTERY, MAKE SURE THAT THE INSTRUMENT HAS BEEN REMOVED
FROM SERVICE UNTIL FULL BATTERY CAPABILITY CAN BE RESTORED.