Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
ii
NON-INVASIVE TOTAL HEMOGLOBIN (SpHb) ACCURACY COMPARED TO INVASIVE
LABORATORY METHODS*
In 492 comparisons of non-invasive total hemoglobin (SpHb) and invasive hemoglobin (tHb)
measurements from a laboratory CO-Oximeter, SpHb accuracy was as follows:
0.90 correlation
0.95 g/dL standard deviation
Below 12 g/dL, 99% of SpHb readings were < 2 g/dL of the laboratory tHb value
At or above 12 g/dL, 94% of SpHb readings were < 2 g/dL of the laboratory value
* Masimo FDA Submission Data
SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES
The Radical-7 Signal Extraction Pulse CO-Oximeter is designed to minimize the
possibility of hazards from errors in the software program by following sound
engineering design processes, Risk Analysis and Software Validation.
The Radical-7 is to be operated by qualifi ed personnel only. This manual,
accessory Directions for Use (DFU), all precautionary information, and
specifi cations should be read before use.
Variation in hemoglobin measurements may be profound and may be affected by
sample type, body positioning, as well as other physiological conditions. As with most
hemoglobin tests, Radical-7 test results should be scrutinized in light of a specific
patient’s condition. Any results exhibiting inconsistency with the patient’s clinical status
should be repeated and/or supplemented with additional test data.
Explosion hazard. Do not use the Radical-7 in the presence of fl ammable
anesthetics or other fl ammable substance in combination with air, oxygen-
enriched environments, or nitrous oxide.
Electric shock hazard. Do not open the Radical-7 cover except to replace the
battery of the Handheld instrument. Only a qualifi ed operator may perform
maintenance procedures specifi cally described in this manual. Refer servicing to
Masimo for repair of this equipment.
High intensity, extreme lights (including pulsating strobe lights) directed on the sensor
may not allow the Radical-7 to obtain readings.
The Radical-7 should be considered an early warning instrument. As a trend
towards patient hypoxemia is indicated, blood samples should be analyzed by
laboratory instruments to completely understand the patient’s condition.
If patient hypoxemia is indicated, blood samples should be analyzed by laboratory
devices to completely understand the patient's condition.
The Radical-7 is NOT intended for use as an apnea monitor.
Pulse rate measurement is based on the optical detection of a peripheral flow pulse
and therefore may not detect certain arrhythmias. The Radical-7 should not be used
as a replacement or substitute for ECG based arrhythmia analysis.
As with all medical equipment, carefully route patient cabling to reduce the
possibility of patient entanglement or strangulation.
Do not place the Radical-7 or accessories in any position that might cause it to
fall on the patient. Do not lift the Pulse CO-Oximeter by the power cord or any
other cable.