Signal Extraction Pulse OximeterOPERATOR’S MANUAL
ix Radical Signal Extraction Pulse Oximeter Operator’s Manual
about this manualThis manual explains how to set up and use the Radical Pulse Oximeter. Important safety information relating to general use of the Ra
overview2 Radical Signal Extraction Pulse Oximeter Operator’s Manual1warnings, cautions and notesPlease read and follow any warnings, cautions and not
product descriptionThe Radical Pulse Oximeter is a noninvasive, arterial oxygen saturation and pulse ratemonitor. Radical can be used as either a Hand
pulse oximetryGENERAL DESCRIPTIONPulse oximetry is a continuous and non-invasive method of measuring the level of arterialoxygen saturation in blood.
FUNCTIONAL VS. FRACTIONAL SATURATIONThe Radical measures and displays functional saturation: the amount of oxygenatedhemoglobin expressed as a percent
introductionRadical provides the functionality of three pulse oximeters in one: Radical is a fully featured Handheld pulse oximeter Radical is a fully
radical handheldThe Handheld Radical provides most of the functionality of the pulse oximeter. All userinput and displays are controlled by this part
system description8 Radical Signal Extraction Pulse Oximeter Operator’s Manual2FastSatMasimo SETPress down the Handheld Release Button and pull the Ha
19Radical Signal Extraction Pulse Oximeter Operator’s Manual 92system descriptionPulseWaveformDisplayTime and DateIndicatorBacklight/ContrastButtonAla
The Radical Operating Instructions are intended to provide the necessary information for properoperation of all models of the Radical pulse oximetry s
system description10 Radical Signal Extraction Pulse Oximeter Operator’s Manual231251 0000 536MEDICAL ELECTRICAL EQUIPMENTUL 2601-1/CAN/CSA C22.2 No.
Radical Signal Extraction Pulse Oximeter Operator’s Manual 11radical standaloneWhen the Radical Handheld is placed into the Docking Station, the Radic
system description12 Radical Signal Extraction Pulse Oximeter Operator’s Manual2STANDALONE FRONT PANELThe following figure and corresponding text revi
Radical Signal Extraction Pulse Oximeter Operator’s Manual 132system descriptionSTANDALONE BACK PANELP2 P3Manufactured by:Masimo CorporationIrvine, CA
14 Radical Signal Extraction Pulse Oximeter Operator’s Manual2SYMBOLSThe following symbols are found on the back of the Radical Docking Station, they
Radical Signal Extraction Pulse Oximeter Operator’s Manual 153introductionBefore the Radical Pulse Oximeter can be used in a clinical setting, it need
16 Radical Signal Extraction Pulse Oximeter Operator’s Manual3 IF THERE IS ANY DOUBT ABOUT THE INTEGRITY OF THE PROTECTIVE EARTHCONDUCTOR ARRANGEMENT,
Radical Signal Extraction Pulse Oximeter Operator’s Manual 173monitor setupThe Radical Pulse Oximeter stores two types of default values: those that R
18 Radical Signal Extraction Pulse Oximeter Operator’s Manual3ALARM VOLUME Set to pre-power down settingPULSE BEEP VOLUME Set to pre-power down settin
Radical Signal Extraction Pulse Oximeter Operator’s Manual 1934. Verify that Radical recognizes the correct cable. The name of the SatShare cable will
SAFETY INFORMATION,WARNINGS, CAUTIONS AND NOTESThe Radical Signal Extraction Pulse Oximeter is designed to minimize the possibility of hazards fromerr
set-up20 Radical Signal Extraction Pulse Oximeter Operator’s Manual3hp vuelink setup1. Select the HP VueLink selection from the Output menu on the Rad
Radical Signal Extraction Pulse Oximeter Operator’s Manual 214introductionTo operate the Masimo SET Radical Pulse Oximeter effectively, the device mus
22 Radical Signal Extraction Pulse Oximeter Operator’s Manual411. Verify that the patient alarms are functional by setting the high and low SpO2andpul
Radical Signal Extraction Pulse Oximeter Operator’s Manual 234operationsuccessful SpO2monitoringThe following general points will aid in ensuring oxim
operation24 Radical Signal Extraction Pulse Oximeter Operator’s Manual4SIGNAL IQThe Radical display provides a visual, indicator of the plethysmogram
Radical Signal Extraction Pulse Oximeter Operator’s Manual 254operationLOW PERFUSIONThe Radical displays an indicator (i.e., “Low Perfusion” message)
26 Radical Signal Extraction Pulse Oximeter Operator’s Manual4touch key control buttons and iconsThe touch key control buttons are the four dark grey
Radical Signal Extraction Pulse Oximeter Operator’s Manual 274navigating the main menuWhen the main menu is accessed, the plethysmograph and Signal IQ
28 Radical Signal Extraction Pulse Oximeter Operator’s Manual4EDITING A PARAMETEROnce a parameter has been selected for editing, a new set of menu sel
Radical Signal Extraction Pulse Oximeter Operator’s Manual 294operationCLOCKABOUTOUTPUTSERVICETime (hour/minute)Time display formatDate (day/month/yea
SAFETY INFORMATION,WARNINGS, CAUTIONS AND NOTES (CONTINUED) The pulse oximeter can be used during defibrillation, but the readings may be inaccurate f
30 Radical Signal Extraction Pulse Oximeter Operator’s Manual4operationalarmsCheck alarm limits each time the pulse oximeter is used to ensure that th
Radical Signal Extraction Pulse Oximeter Operator’s Manual 314operationThis menu allows the user to set the alarm silence period. Analarm is silenced
32 Radical Signal Extraction Pulse Oximeter Operator’s Manual4operationMENU ITEMS DESCRIPTIONThree views are available: Pleth and Signal IQ, Numbers a
Radical Signal Extraction Pulse Oximeter Operator’s Manual 334operationgeneralAVERAGING TIME The signal averaging time of this device can be set to: 2
34 Radical Signal Extraction Pulse Oximeter Operator’s Manual4operationaboutThis displays the copyright and software versions of the Handheld and Dock
Radical Signal Extraction Pulse Oximeter Operator’s Manual 354operationCAUTION: TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANYEQUIPMENT TO
36 Radical Signal Extraction Pulse Oximeter Operator’s Manual4operationtrend displayOnce the Trend Display touch key icon is selected the trend data i
Radical Signal Extraction Pulse Oximeter Operator’s Manual 374operationThe top line on the trend display shows the time scale of the trend graph, foll
38 Radical Signal Extraction Pulse Oximeter Operator’s Manual4operationNAVIGATING THE TREND DISPLAYIn the Trend Display view there are a total of 10
Radical Signal Extraction Pulse Oximeter Operator’s Manual 394operationTREND SETUPThis menu allows the user to set the default trend settings and to c
table of contentsSECTION 1 / OVERVIEWAbout this Manual ...
40 Radical Signal Extraction Pulse Oximeter Operator’s Manual4TREND PERIODThe Trend period setting determines how often a set of SpO2and pulse rate da
Radical Signal Extraction Pulse Oximeter Operator’s Manual 414 The user can access the menu system. If the SatShare cable is connected to the Radical
42 Radical Signal Extraction Pulse Oximeter Operator’s Manual4home mode operationThe Radical can be placed into the Home Mode to protect unqualified u
Radical Signal Extraction Pulse Oximeter Operator’s Manual 435introductionThe following sections outline system alarms, messages displayed on the LCD
44 Radical Signal Extraction Pulse Oximeter Operator’s Manual5alarms/messagesalarm operationFor averaging modes of less than 14 seconds, the Radical i
Radical Signal Extraction Pulse Oximeter Operator’s Manual 455alarms/messagessystem messagesThe following chart alphabetically lists all system messag
46 Radical Signal Extraction Pulse Oximeter Operator’s Manual5alarms/messages* The SERVICE REQUIRED message fills the entire display.NO SENSOR Sensor
Radical Signal Extraction Pulse Oximeter Operator’s Manual 476troubleshootingtroubleshootingThe following chart describes what to do if the Radical sy
48 Radical Signal Extraction Pulse Oximeter Operator’s Manual7specificationsradical specificationsPERFORMANCEmeasurement rangeSpO2: 1-100%Pulse Rate:
Radical Signal Extraction Pulse Oximeter Operator’s Manual 497specificationsENVIRONMENTALOperating Temperature: 41°F to 104°F (5°C to 40°C)Storage Tem
SECTION 4 / OPERATIONIntroduction ...
50 Radical Signal Extraction Pulse Oximeter Operator’s Manual7specificationscomplianceEMC Compliance: EN60601-1-2, Class BEquipment Classification: IE
Radical Signal Extraction Pulse Oximeter Operator’s Manual 517specificationsPARAMETER SETTINGBAUD RATE 9600 Baud bi-directionalNUMBER OF BITS PER CHAR
52 Radical Signal Extraction Pulse Oximeter Operator’s Manual7specificationsanalog output / nurse call specificationsThe Analog Out and Nurse Call are
Radical Signal Extraction Pulse Oximeter Operator’s Manual 537specificationsANALOG OUTPUTRadical can interface with various analog recording devices a
54 Radical Signal Extraction Pulse Oximeter Operator’s Manual8sensor/patient cablesintroductionThis section covers the use and cleaning of Masimo LNOP
Radical Signal Extraction Pulse Oximeter Operator’s Manual 558sensor/patient cablesCLEANING AND REUSE OF MASIMO LNOP SENSORSReusable sensors can be cl
56 Radical Signal Extraction Pulse Oximeter Operator’s Manual8sensors/patient cablesmasimo SET patient cablesReusable patient cables of various length
Radical Signal Extraction Pulse Oximeter Operator’s Manual 579service/maintenanceintroductionThis chapter covers how to test the operation of the Radi
58 Radical Signal Extraction Pulse Oximeter Operator’s Manual9service/maintenancebattery operation and maintenanceThe Radical Handheld includes a 1.5
Radical Signal Extraction Pulse Oximeter Operator’s Manual 599service/maintenanceCAUTION: IF THE RADICAL HAS NOT BEEN CHARGED FOR 1 MONTH OR MORE, THE
Home Use...33Interface Alarms ...
60 Radical Signal Extraction Pulse Oximeter Operator’s Manual9service/maintenanceWARNING: THE DOCKING STATION BATTERY SHOULD BE INSTALLED AND/ORREMOVE
Radical Signal Extraction Pulse Oximeter Operator’s Manual 619service/maintenanceREPLACING THE FUSESShould a power problem blow one or both of the fu
62 Radical Signal Extraction Pulse Oximeter Operator’s Manual9service/maintenancePower-On Self-Test:1. Connect the monitor to AC power and verify that
Radical Signal Extraction Pulse Oximeter Operator’s Manual 639service/maintenanceTesting with Masimo SET Tester (Optional):1. Turn the Radical off and
64 Radical Signal Extraction Pulse Oximeter Operator’s Manual9service/maintenanceAnalog Output Test1. Disconnect all patient cables and sensors from t
Radical Signal Extraction Pulse Oximeter Operator’s Manual 659service/maintenanceservice and repairREPAIR POLICYMasimo or an authorized Service Depart
service/maintenance66 Radical Signal Extraction Pulse Oximeter Operator’s Manual9warrantyMasimo warrants to the initial purchaser that each new pulse
Radical Signal Extraction Pulse Oximeter Operator’s Manual 6710accessoriesACCESSORIES PART NUMBERRADICAL HANDHELD 1309RADICAL DOCKING STATION RDS-1 13
accessories68 Radical Signal Extraction Pulse Oximeter Operator’s Manual10ACCESSORIES PART NUMBERSATSHARE CABLE-DO01 1324SATSHARE CABLE-DO02 1326SATSH
Backlight / Contrast Operation.......40SatShare Operation
Radical Signal Extraction Pulse Oximeter Operator’s Manual viiitable of contentsSECTION 9 / SERVICE AND MAINTENANCEIntroduction ......
Commentaires sur ces manuels