52 Radical Signal Extraction Pulse Oximeter Operator’s Manual
7
specifications
analog output / nurse call specifications
The Analog Out and Nurse Call are features accessible on the same female high density
DB-15 connector.
NOTE: The Radical analog output / nurse call interface is only available when the Radical
Handheld is properly attached to the Radical Docking Station.
NOTE: The analog output/nurse call interface is no available in all version of the Docking
Station.
The following table shows the pinout of the analog output and nurse call.
PIN SIGNAL NAME
1 +5V (60mA max.)
2 Ground
3 Ground
4 Ground
5 Ground
6 Nurse Call (Normally Open)
7 Nurse Call (Normally Closed)
8 Ground
9 Analog Out - %SpO
2
10 Ground
11 Ground
12 Nurse Call Common
13 Ground
14 Ground
15 Analog Out Pulse Rate
WARNING: ALL EXTERNAL DEVICE CONNECTIONS TO THE RS-232 SERIAL
PORT MUST BE IEC-60950 COMPLIANT.