Radical Signal Extraction Pulse Oximeter Operator’s Manual 55
8
sensor/patient cables
CLEANING AND REUSE OF MASIMO LNOP SENSORS
Reusable sensors can be cleaned per the following procedure:
■
Remove the sensor from the patient.
■
Disconnect the sensor from the monitor.
■
Wipe the entire sensor clean with a 70% isopropyl alcohol pad.
■
Allow the sensor to air dry before returning it to operation.
REATTACHMENT OF SINGLE USE ADHESIVE SENSORS
■
LNOP single use sensors may be reapplied to the same patient if the emitter and
detector windows are clear and the adhesive still adheres to the skin.
■
The adhesive can be partially rejuvenated by wiping with a 70% isopropyl alcohol
wipe and allowing the sensor to thoroughly air dry prior to replacement on the
patient.
NOTE: If the sensor fails to track the pulse consistently, the sensors may be incorrectly
positioned. Reposition the sensor or choose a different monitoring site.
SENSOR USAGE PATIENT WEIGHT
LNOP ADT Single Use Adults > 30 kg
LNOP PDT Single Use Pediatrics and Slender Adults >10 kg and < 50 kg
LNOP NEO Single Use Neonate < 10 kg
LNOP NEO PT Single Use Neonate < 1 kg, or with poor skin integrity
LNOP DCI Reusable Adults and Pediatrics > 30 kg
LNOP DCIP Reusable Pediatrics >10 kg and <50 kg
LNOP DCSC Reusable Adult and Pediatrics >30 kg, for spot check applications
WARNING: TO AVOID CROSS CONTAMINATION ONLY USE MASIMO LNOP
SINGLE USE SENSORS ON THE SAME PATIENT.
CAUTIONS:
■
DO NOT REPROCESS ANY LNOP SINGLE USE SENSORS.
■
DO NOT SOAK OR IMMERSE THE SENSOR IN ANY LIQUID SOLUTION.
DO NOT STERILIZE ANY MASIMO SENSOR BY IRRADIATION, STEAM,
OR ETHYLENE OXIDE.