Radical Signal Extraction Pulse Oximeter Operator’s Manual 47
6
troubleshooting
troubleshooting
The following chart describes what to do if the Radical system does not operate properly
or fails.
PROBLEM POSSIBLE CAUSE(S) RECOMMENDATION
UNIT DOES NOT One or both of the fuses Replace the fuses.
POWER ON. have blown.
UNIT POWERS ON The viewing contrast Use the Backlight/Contrast button to adjust
BUT THE GRAPHIC is not correct. the viewing angle. If the condition
DISPLAY IS BLANK. persists, the unit requires service.
CONTINUOUS Internal failure. Unit requires service. Press the Alarm
SPEAKER TONE. Silence button to silence the alarm. If
alarm continues to sound, power down
unit and remove Handheld battery if
necessary.
BUTTONS DON’T Internal failure. Unit requires service.
WORK WHEN
PRESSED
The following chart describes what to do when encountering common problems:
PROBLEM POSSIBLE CAUSE(S) RECOMMENDATION
HANDHELD BATTERY AC power cable may be Restore power to the device.
DOES NOT CHARGE. disconnected.
AC power switch is turned
off.
PRINT FUNCTION Wrong serial cable is used. Make sure a null modem cable is
DOES NOT WORK. used.