Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 7-17-1
7
specifications
Radical-7 specifications
PERFORMANCE
Measurement Range
Oxygen Saturation (%SpO
2
): 0-100%
Carboxyhemoglobin Saturation (%SpCO): 0-99%
Methemoglobin Saturation (%SpMet): 0-99.9%
Pulse Rate: 25-240 beats per minute (bpm)
Perfusion Index: 0.02% - 20%
Response Time: <1 second delay
ACCURACY
Saturation 70% to 100%
No Motion
1
Adults, Pediatrics ±2 digits
Neonate
2
±3 digits
Motion
3
Adults, Pediatrics
±3 digits
Neonate ±3 digits
Low Perfusion
4
Adults, Pediatrics ±2 digits
Neonate ±3 digits
Carboxyhemoglobin Saturation Accuracy (%SpCO)
5
1% - 40% ±3 digits
Methemoglobin Saturation Accuracy (%SpMet)
5
1% - 15% ±1 digit
Pulse Rate Accuracy
10
Pulse rate:
25-240 bpm
No Motion
1
Adults, Pediatrics, Neonate ±3 digits
Motion
3
Adults, Pediatrics, Neonate ±5 digits
Low Perfusion
4
Adults, Pediatrics, Neonate ±3 digits
Resolution
Saturation (%SpO
2
)1%
Carboxyhemoglobin saturation (%SpCO), digital display 1%
Methemoglobin saturation (%SpMet), digital display .1%
Pulse Rate (bpm) 1 bpm
ELECTRICAL
Standalone
AC Power requirements: 100-240 VAC, 47-63 Hz
Power consumption: 55 VA
Fuses: 1 Amp, Fast Acting, Metric, (5x20mm), 250V
Batteries
Handheld:
Type: NiMH
Capacity: 4 hours
6
Charging time: 3 hours