2-4 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 2-5
system description
2
21
SIGNAL IQ
The Signal IQ shows the acquired signal quality and the
timing of the pulse. A tall vertical line indicates a high qual-
ity signal, while a small vertical line indicates a low quality
signal. The Signal IQ may will be displayed as a single,
pulsating bar in the Numbers mode.
22
BRIGHTNESS
LEVEL
The Brightness level icon displays when utilizing the
Backlight/Contrast button.
23
3.00
PI
PERFUSION
INDEX
The Perfusion Index indicates numerically the percentage of
pulsatile signal to non-pulsatile signal (pulse strength).
24
APOD
SENSITIVITY
The MAX or APOD™ sensitivity icon is shown on the
Radical-7 display to indicate if the Radical-7 is set to oper-
ate in Normal sensitivity, Maximum sensitivity or Adaptive
Probe Off Detection mode
25
140
50
PULSE RATE
ALARM LIMITS
DISPLAY
The Pulse Rate Alarm Limits Display shows the upper
and lower pulse rate alarm limits. When an alarm limit is
reached or exceeded, the pulse rate value and the violated
limit flashes.
26
FastSat
FAST S AT
The FastSat
®
label is shown on the Radical-7 display when-
ever the Radical-7 is set to operate in the FastSat mode.
27
76
PULSE RATE
The Pulse Rate Measurement Display shows the patient’s
pulse rate in beats per minute. The upper and lower pulse
rate alarm limits are also displayed next to the pulse rate
measurement. The pulse rate is calculated and the display
is updated at a frequency of once per second.
28
Philips VueLink
SERIAL
OUTPUT MODE
The Serial Output Mode displays the selected output interface
when interfaced with the Docking Station. and connected to the
serial port with the selected Satshare cable.
* For instruments that include SpMet and SpCO parameters: the SpMet and SpCO
parameters will be grayed out in the display screen if a patient cable is not attached or a
non-Rainbow sensor is being used.
NOTE: The Pulse CO-Oximeter display view will be different without the optional SpCO
and SpMet options.
system description
2-4
2
10
50
5
SpCO ALARM
LIMITS DISPLAY
The SpCO Alarm Limits Display shows the upper and lower
alarm limits. When the measured value is outside of the
alarm limits, the SpCO measurement display flashes and an
alarm will sound
11
PULSE
WAVEFORM
DISPLAY
The Pulse Waveform Display shows the acquired plethysmo-
graph waveform. The pleth waveform is scaled with signal
strength. Signal strength is defined as the relation of arterial
pulsatile signal to the non-pulsatile signal component.
12
BATTERY
STATUS
INDICATORS
The Battery Status Indicators show the capacity of the
Radical-7 Handheld and optional Docking Station batter-
ies. The indicator flashes when less than 15 minutes of bat-
tery life is left and the battery needs to be recharged. The
Docking Station Battery Status indicator is not shown when
the optional Docking Station battery is not present.
13
TOUCH KEY
CONTROL
BUTTONS
Press a Touch Key Control Button to select the correspond-
ing touch key icon. See Section 4, Touch Key Control
Buttons and Icons for more details.
14
BACKLIGHT
BUTTON
Press the Backlight Button to change the illumination level
of the backlight. With the AC line power connected, four
levels of illumination are available (in addition to the no
illumination level). In the Handheld mode, three levels of
illumination are available (again in addition to the no illumi-
nation level). Use the lowest illumination for most efficient
battery usage.
15
ALARM SILENCE
BUTTON
Press the Alarm Silence Button to temporarily silence
patient and low battery alarms. Press the Alarm Silence
Button when the SENSOR OFF message is flashing (i.e.
the sensor is removed from the patient) to acknowledge
the end of monitoring. In this state, all further alarms are
suspended until the Pulse CO-Oximeter starts measuring
SpO
2
, SpMet, SpCO and pulse rate again.
Note: System failure alarms can be silenced by pressing
the Power/Standby or Alarm Silence Button. If the Power/
Standby Button does not silence the system fault alarm,
press the Alarm Silence Button
16
SPEAKER
The speaker indicates audio alarms. Care should be taken
not to cover the speaker and muffle the audible alarm
volume.
17
PATIENT CABLE
CONNECTOR
Connect a patient cable or a direct cable sensor into the
Handheld Radical-7 by plugging the cable into the Patient
Cable Connector. Use only Masimo compatible sensors
and cables with this oximeter. See Section 8, Sensors and
Patient Cables for more details.
18
POWER/ON/OFF
BUTTON
Press the Power/On/Off Button to turn the instrument on.
Press, hold the button for more than 2 seconds and then
release the button to turn the instrument off.
19
APOD
T
REND
GRAPH
MAX
TOUCH KEY
ICONS
The Touch Key Icons indicate the software menu items that
can be selected through the Touch Key Control Buttons.
Pressing a Touch Key Control Button next to an icon selects
the option.
20
10/11/06 12:23:11am
TIME AND DATE
INDICATOR
The Time and Date Indicator displays the current time and
date. The time is displayed in 12 or 24 hour format. The date
is displayed in dd/mm/yy or mm/dd/yy format. Select the
date and time display formats in the Clock menu.