3-4 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 3-5
OPTION DEFAULT SETTING
TREND PERIOD
Set to 1 Hr.
LOW SpO
2
ALARM LIMIT
Set to 0
SpMet HIGH/LOW ALARM LIMIT
Set to High: 3.0/Low: ---
SpCO HIGH/LOW ALARM LIMIT
Set to High: 45/Low: ---
SMART TONE
Set to Yes
ADULT/NEO (HOSPITAL DEFINED) SETTINGS
The following table outlines settings that may be defined by the hospital. If enabled (via a
password protected screen, see Section 4 Password Operation), these settings will return
to pre defined values after a power cycle. This mode is indicated by “Mode Adult” or “Mode
Neo” on the Alarms menu. See Section 4, Operation for details on enabling and setting
these parameters.
OPTION
CUSTOM SETTING
(DEFAULT)
ADULT / NEO SETTING
(PRE DEFINED)
SPO
2
HIGH/LOW ALARM
LIMIT
Set to High: ---/Low: 90
Same settings
PULSE RATE HIGH/LOW
ALARM LIMIT
Set to High: 140/Low: 50 BPM
SpMet HIGH/LOW ALARM
LIMIT
Set to High: 3.0/Low: ---
SpCO HIGH/LOW ALARM
LIMIT
Set to High: 45/Low: ---
ALARM SILENCE
Set to 120 seconds
ALARM VOLUME
Set to 3
ALARM DELAY
Set to 5
Satshare Setup
The Radical-7 Pulse CO-Oximeter has been proven to be accurate during patient motion
and low perfusion conditions. Saturation and pulse rate values from the Radical-7 Pulse
CO-Oximeter may be displayed on a multiparameter monitor through the SatShare fea-
ture.
The SatShare feature provides an ideal, simulated waveform corresponding to the mea-
sured saturation and pulse rate values determined by the Masimo SET technology. This
waveform may be used to display these values on multiparameter monitors through the
multiparameter monitor’s oximetry sensor or input connector.
It is recommended that the Radical-7 Pulse CO-Oximeter is positioned close to the mul-
tiparameter monitor with the Radical-7 Pulse CO-Oximeter screen visibly displaying the
plethysmographic waveform and the saturation and pulse rate measurements.
CAUTION: SIMULTANEOUS USE OF SATHSARE AND SERIAL PORT IS NOT
SUPPORTED.
SATSHARE SETUP
1. Select the SatShare cable that is appropriate for the multiparameter monitor that
is being connected. Check the Masimo web site at www.masimo.com for the lat-
est list of available SatShare cables and validated instruments.
2. Connect the labeled end of the cable to the SatShare Cable Connector port on
the back of the Docking Station. Tighten the connector screws for a secure con-
nection.
3. Connect the other end of the SatShare cable either to the sensor connector of
the multiparameter monitor’s SpO
2
cable or directly to the SpO
2
connector on the
monitor.
4. Verify that the Radical-7 Pulse CO-Oximeter recognizes the correct cable. The
name of the SatShare cable will be displayed on the LCD screen when the
SatShare mode is functional.
5. Set the multiparameter monitor’s high and low saturation and pulse rate alarm
limits as appropriate.
6. Set the multiparameter monitor’s averaging time to the lowest setting (i.e. fastest
response). The Radical-7 Pulse CO-Oximeter’s ideal waveform necessitates the
need for additional averaging by the monitor. If the multiparameter monitor’s aver-
aging time is not changed, the time to display physiological changes in saturation
on the monitor will be increased with SatShare. However, the delay can be mini-
mized by reducing the multiparameter monitor’s averaging time.
7. While in the SatShare mode, if there are any significant discrepancies between
the readings from the Radical-7 Pulse CO-Oximeter and those on the monitor dis-
playing the values obtained from SatShare, the values reported by the Radical-7
Pulse CO-Oximeter are to be considered the correct values.
8. To use the Radical-7 Pulse CO-Oximeter with SatShare while it is not connected
to AC power, set the Power Save parameter in the General menu to “No” and
refer to Section 4-Operation. Please note that if the Radical-7 Pulse CO-Oximeter
is used in this mode, the length of time the Radical-7 Pulse CO-Oximeter can
operate on battery power will be significantly diminished.
setup
3
3
setup
S
ig
n
a
l E
x
tra
cti
o
n
P
u
ls
e
C
O
-O
x
im
e
te
r
M
asim
o
r
a
i
n
b
o
w
SET
9
9
8
5
S
p
0
2
14
0
5
0
B
P
M
P
I
S
p
M
et
S
pC
O
3
.0
0
.3
5
0
5
10/11/06
12:23:11am
Vue de la page 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 54 55

Commentaires sur ces manuels

Pas de commentaire