1-6 Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual
1
Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual 1-7
1
are abnormal (less than 90% for arterial oxygen saturation, and greater than 1% for
methemoglobin concentration). As blood gas samples are usually taken over a period
of 20 seconds (the time it takes to draw the blood) a meaningful comparison can only be
achieved if the core oxygen saturation and carboxyhemoglobin concentration of the patient
are stable and not changing over the period of time that the blood gas sample is taken.
MASIMO SET SIGNAL EXTRACTION TECHNOLOGY FOR SpO
2
MEASUREMENTS
Masimo Signal Extraction Technology’s signal processing differs from conventional
pulse oximeters. Conventional pulse oximeters assume that arterial blood is the only
blood moving (pulsating) in the measurement site. During patient motion, however, the
non-arterial blood also moves, causing conventional pulse oximeters to read low values,
because they cannot distinguish between the arterial and venous blood movement
(sometimes referred to as noise). Masimo SET pulse oximetry utilizes parallel engines and
adaptive digital filtering. Adaptive filters are powerful because they are able to adapt to the
varying physiologic signals and/or noise and separate them by looking at the whole signal
and breaking it down to its fundamental components. The Masimo SET signal processing
algorithm, Discrete Saturation Transform
TM
(DST), reliably identifies the noise, isolates it
and, using adaptive filters, cancels it. It then reports the true arterial oxygen saturation for
display on the monitor.
SpCO MEASUREMENTS DURING PATIENT MOTION
Rad-57 displays measurements of SpCO during patient motion. However, because of the
changes in the physiological parameters such as blood volume, arterial-venous coupling,
etc. that occur during patient motion, the accuracy of such measurements is not reliable.
overview
Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
1
Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
1
overview
Digitized,
Filtered &
Normalized
R/IR
R/IR
(Conventional
Pulse Oximetry)
Adaptive
Filter with
DST
TM
Adaptive
Filter with
SST
TM
FST
TM
Proprietary
Algorithm 4
Saturation
Post
Processor
Confidence
Based
Arbitrator
MEAS
CONF
MEAS
CONF
MEAS
CONF
MEAS
CONF
MEAS
CONF
MASIMO SET PARALLEL ENGINES
MASIMO SET DST