Masimo Radical Signal Extraction Manuel de l'opérateur Page 15

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 30
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 14
4-6 Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual
4
Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual 4-7
4
operation
NUMERIC DISPLAY - SpCO
A stable SpCO reading is associated with correct sensor placement, small physiological
changes during the measurement and acceptable levels of arterial perfusion in the patient's
fingertip (measurement site). Physiological changes at the measurement site are mainly
caused by fluctuations in the oxygen saturation, blood concentration and perfusion.
Inaccurate measurements may be caused by:
Significant levels of methemoglobin.
Intravascular dyes such as indocyanine green or methylene blue.
Abnormal hemoglobin levels.
Abnormally low arterial perfusion.
NUMERIC DISPLAY - PULSE RATE
The Pulse Rate displayed on the Rad-57 may differ slightly from the heart rate displayed
on ECG monitors due to differences in averaging times. There may also be a discrepancy
between cardiac electrical activity and peripheral arterial pulsation. Significant differences
may indicate a problem with the Signal IQ due to physiological changes in the patient or
one of the instruments or application of the sensor or patient cable. The pulsations from
intra-aortic balloon support can be additive to the pulse rate displayed on the Pulse CO-
Oximeter .
LOW SIGNAL IQ (LOW SIQ)
The Rad-57 display provides a visual indicator Signal IQ and an alert when the displayed
SpO
2
values are not based on adequate Signal IQ. The Signal IQ indicator displayed on
the Rad-57 as “Low SIQ”.
The Low SIQ indicator flashes when the SpO
2
measurement may be compromised. When
the Low SIQ indicator is flashing, proceed with caution and do the following:
Assess the patient.
Check the sensor and ensure proper sensor application. The sensor must be
well secured to the site for the Rad-57 to maintain accurate readings. Also,
misalignment of the sensor’s emitter and detector can result in smaller signals.
Determine if an extreme change in the patient’s physiology and blood flow at the
monitoring site occurred, (e.g. an inflated blood pressure cuff, a squeezing motion,
sampling of an arterial blood specimen from the hand containing the pulse oximetry
sensor, severe hypotension, peripheral vasoconstriction in response to hypothermia,
medications, or a spell of Raynaud’s syndrome.)
With neonates or infants, check that the peripheral blood flow to the sensor site
is not interrupted. For example, as may occur while lifting or crossing their legs,
during a diaper change.
operation
After performing the above, if the “Low SIQ” indication occurs frequently or continuously,
obtaining an arterial blood specimen for CO-Oximetry analysis may be considered to verify
the oxygen saturation value.
LOW PERFUSION / PERFUSION INDEX (PI)
Perfusion Index, or PI, is a relative assessment of the perfusion at the monitoring site. PI
is displayed on a 10 segment LED bar, ranging from <1% (very weak perfusion) to >5%
(strong perfusion). The lower two segments of the bar will turn red when the amplitude
of the arterial pulsations are very low (weak perfusion). The highest LED will remain lit
continuously to allow a PI level to be viewed.
The PI is shown as a "bouncing bar" indicator, where the peak of the bar coincides with the
peak of an arterial pulsation. Even with a plethysmographic waveform obscured by artifact,
the Rad-57 locates the arterial pulsation. The pulse tone (when enabled) coincides with
the peak of the PI bar.
CAUTION: IF THE LOW PERFUSION INDICATION IS FREQUENTLY DISPLAYED, FIND
A BETTER PERFUSED MONITORING SITE. IN THE INTERIM, ASSESS THE PATIENT
AND, IF INDICATED, VERIFY OXYGENATION STATUS THROUGH OTHER MEANS.
ACTIONS TO BE TAKEN
If the SpO
2
readings show significant differences, do the following:
Make sure the emitter and photodetector are aligned directly opposite each other.
Select a site where the distance between the emitter and photodetector is minimized.
Wipe the sensor site with a 70% isopropyl alcohol pad or rubefacient cream (10-
30% methyl salicylate and 2-10% menthol) for 20-30 seconds. Strong vasodilator
creams, such as nitroglycerin paste, are not recommended.
If possible, remove electrical noise sources such as electrosurgical units or other
electrical/electronic equipment.
If artificial nails or excessive fingernail polish are present, select another site or
remove the polish/artificial nails.
If possible, ensure that the sensor is placed in a location with low ambient light.
CAUTION: IF ANY MEASUREMENT SEEMS QUESTIONABLE, FIRST CHECK THE
PATIENT’S VITAL SIGNS BY ALTERNATE MEANS AND THEN CHECK THE PULSE CO-
OXIMETER FOR PROPER FUNCTIONING
Vue de la page 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 29 30

Commentaires sur ces manuels

Pas de commentaire