1-6 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 1-7
1
overviewoverview
1
FUNCTIONAL SATURATION
The Radical-7 is calibrated to measure and display functional saturation (SpO
2
): the
amount of oxyhemoglobin expressed as a percentage of the hemoglobin that is available
to transport oxygen. Note that carboxyhemoglobin and methemoglobin are not capable
of transporting oxygen, but is recognized as oxygenated hemoglobin by conventional
pulse oximetry.
MEASURED VS. CALCULATED SATURATION
SpO
2
, SpMet and SpCO measurements that can be obtained from the Radical-7 are
commonly compared to invasive measurements obtained from blood gas samples. When
comparing invasive and noninvasive measurements and interpreting values, caution
should be used, as the calculated values obtained from the blood gas sample may differ
from the SpO
2
, SpMet and SpCO measurements of the Pulse CO-Oximeter. In the case
of SpO
2
, different results are usually obtained from the arterial blood gas sample if the
calculated measurement is not appropriately corrected for the effects of variables that
shift the relationship between the partial pressure of oxygen (PO2) and saturation, such
as: pH, temperature, the partial pressure of carbon dioxide (PCO2), 2,3-DPG, and fetal
hemoglobin. In the case of SpMet and SpCO, in addition to the effects of temperature
and pH, different results are also expected if concentration of methemoglobin in the blood
gas sample are abnormal (less than 90% for arterial oxygen saturation, and greater than
2% for methemoglobin concentration). As blood gas samples are usually taken over a
period of 20 seconds (the time it takes to draw the blood) a meaningful comparison
can only be achieved if the oxygen saturation, carboxyhemoglobin and methemoglobin
concentration of the patient are stable and not changing over the period of time that the
blood gas sample is taken.
MASIMO SET SIGNAL EXTRACTION TECHNOLOGY FOR SpO
2
MEASUREMENTS
Masimo Signal Extraction Technology’s signal processing differs from conventional pulse
oximeters. Conventional pulse oximeters assume that arterial blood is the only blood mov-
ing (pulsating) in the measurement site. During patient motion, however, the non-arterial
blood also moves, causing conventional pulse oximeters to read low values, because
they cannot distinguish between the arterial and venous blood movement (sometimes
referred to as noise). Masimo SET pulse oximetry utilizes parallel engines and adaptive
digital fi ltering. Adaptive fi lters are powerful because they are able to adapt to the varying
physiologic signals and/or noise and separate them by looking at the whole signal and
breaking it down to its fundamental components. The Masimo SET signal processing
algorithm, Discrete Saturation Transform
®
(DST
®
) reliably identifi es the noise, isolates it
and, using adaptive fi lters, cancels it. It then reports the true arterial oxygen saturation
for display on the monitor.
SpMet AND SpCO MEASUREMENTS DURING PATIENT MOTION
The Radical-7 displays measurements of SpMet and SpCO during patient motion.
However, because of the changes in the physiological parameters such as blood volume,
arterial-venous coupling, etc. that occur during patient motion, the accuracy of such mea-
surements may not be readable during excessive motion.
MASIMO RAINBOW SET PARALLEL ENGINES
MASIMO SET DST
®
Intermediate Calculations
Confidence
Intermediate Calculations
Confidence
Intermediate Calculations
Confidence
Intermediate Calculations
Confidence
Intermediate Calculations
Confidence
Intermediate Calculations
Confidence
Intermediate Calculations
Confidence
Pre-Processing &
Data Reduction
Digitized, Filtered &
Normalized
LED
Wavelengths
R/IR
(Conventional
Pulse Oximetry)
DST
Adaptive
Filter
SST ™
®
FST:
Proprietary
Algorithm 3
FST:
Proprietary
Algorithm 4
Rainbow
Specific
Algorithms
Pulse
Rate
Algorithm
Confidence
Based
Arbitrator
Post
Processor
%SpO
2
%SpCO
%SpMet
PI
PR
This fi gure is for conceptual purposes only.