Masimo Radical Signal Extraction Manuel de l'opérateur

Naviguer en ligne ou télécharger Manuel de l'opérateur pour Détecteurs de monoxyde de carbone (CO) Masimo Radical Signal Extraction. Masimo Radical Signal Extraction Operator`s manual [en] Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Pulse Oximeter

Rad-8®Pulse OximeterSignal Extraction Pulse Oximeter OPERATOR’S MANUAL

Page 2

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 1-11overviewAbout This ManualThis manual explains how to set up and use the Rad-8 Pulse Oxime

Page 3

1-2 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual1overviewWarnings, cautions and notesPlease read and follow any warnings, cautions and not

Page 4 - i

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 1-31overviewProduct DescriptionThe Rad-8 is a noninvasive, arterial oxygen saturation and pul

Page 5

1-4 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual1overview1Pulse OximetrySpO2 GENERAL DESCRIPTIONPulse Oximetry is a continuous and noninva

Page 6

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 1-51overviewPRINCIPLE OF OPERATION (CONTINUED)The Rad-8 Pulse Oximeter uses a two-wavelength

Page 7

1-6 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual1overview1SIGNAL EXTRACTION TECHNOLOGY (SET) FOR SpO2 MEASUREMENTSMasimo Signal Extraction

Page 8 - v

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 1-711overviewMASIMO SET DST

Page 10

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 2-122system descriptionIntroductionThe Rad-8 is designed for ease of operation. All pulse oxi

Page 11

2-2 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual2system descriptionRAD-8 FRONT PANEL (VERTICAL MODEL)888888®888888RAD-8 FRONT PANEL (HORIZ

Page 12

Rad-8Pulse OximeterSignal Extraction Pulse OximeterOPERATOR’S MANUAL

Page 13 - Absorption Spectra

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 2-32system descriptionFront Panel Controls/IndicatorsCONTROL / INDICATORDESCRIPTION1AC Power

Page 14

2-4 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual2system descriptionFront Panel Controls/Indicators continuedCONTROL / INDICATORDESCRIPTION

Page 15

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 2-52SYMBOLSThe following symbols are found on the pulse oximeter or packaging and are defined

Page 17

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 3-13setupIntroductionBefore the pulse oximeter can be used in a clinical setting, it needs to

Page 18

3-2 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual3setupINITIAL BATTERY CHARGINGBefore use, the battery needs to be fully charged.To charge

Page 19

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-14operationIntroductionTo operate the Rad-8 system effectively, the device must be set up c

Page 20 - DESCRIPTION

4-2 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4operationGENERAL SETUP AND USE (CONTINUED)13. Verify the sensor alarms are functional.

Page 21 - RAD-8 REAR PANEL

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-34operationFACTORY DEFAULT AND USER CONFIGURABLE SETTINGSOPTIONFACTORY DEFAULT SETTINGSUSER

Page 22 - 5%-95% RH

4-4 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4operationSuccessful MonitoringThe following general points will aid in ensuring monitorin

Page 24

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-54operationNUMERIC DISPLAY - PULSE RATEThe Pulse Rate displayed on the Rad-8 Pulse Oximeter

Page 25 - HUMIDITY

4-6 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4operationACTIONS TO BE TAKENIf the SpO2 readings show significant differences, do the fo

Page 26

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-74operationSENSOR PLACEMENTIf the SpO2, readings are questionable or unavailable, do the fo

Page 27

4-8 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4operationLOW BATTERY AUDIBLE ALARMIf a low battery condition occurs during patient monito

Page 28

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-94operationNormal patient monitoringFRONT PANEL CONTROL OPERATIONSETUP MENUThis section giv

Page 29

4-10 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4SETUP MENU LEVEL 1 (CONTINUED)LED BRIGHTNESSThe Display screen and all active LED indica

Page 30

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-114operationSETUP MENU LEVEL 2 (CONTINUED)ALARM SILENCE BUTTONS

Page 31

4-12 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4SETUP MENU LEVEL 2 (CONTINUED) CLEAR TRENDThe Rad-8 only stores data in the trend memory

Page 32

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-134operationSENSITIVITY BUTTONS SETTIN

Page 33

4-14 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4SETUP MENU LEVEL 3 (ENTER BUTTON + DOWN BUTTON MENU SETTINGS)Setup Menu Level 3 contains

Page 34

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual iThe Rad-8 Operating Instructions provide the necessary information for proper o

Page 35

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-154operationSETUP MENU LEVEL 3 - (CONTINUED)RAPID DESAT LIMITThe Rapid Desat Limit is desig

Page 36 - ALARM DELAY

4-16 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4SETUP MENU LEVEL 3 - (CONTINUED)DEFAULT SETTINGS BUTTONS

Page 37

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-174operationSETUP MENU LEVEL 3 - (CONTINUED)YEAR BUTTONS

Page 38 - SENSITIVITY

4-18 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4SETUP MENU LEVEL 3 (CONTINUED)DAY BUTTONS

Page 39 - + DOWN BUTTON MENU SETTINGS)

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-194operationSETUP MENU LEVEL 3 (CONTINUED)MINUTE BUTTONS

Page 40 - ALARM ON/OFF

4-20 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4SETUP MENU LEVEL 3 (CONTINUED)SERIAL OUTPUT BUTTONS

Page 41

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-214operationNURSE CALL BUTTONS SETTINGS I

Page 42

4-22 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4SETUP MENU LEVEL 4 (ENTER BUTTON + UP BUTTON MENU SETTINGS)Level 4 parameter/measurement

Page 43

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-234operationTrend Setup and UseINTRODUCTIONThe Rad-8 can store up to 72 hours of trend data

Page 44 - 3 (CONTINUED)

4-24 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual4TREND DATA FORMATAfter a successful download of the trend data, a .out file will be crea

Page 45

Rad-8 Signal Extraction Pulse Oximeter Operator’s ManualiiSAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTESThe Rad-8 Pulse Oximeter is designed to mini

Page 46

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-254operationHOME MODE OPERATIONThe device can be placed into the Home Mode to protect unqua

Page 48

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 5-15messagesAlarm identificationThe device visually and audibly indicates alarm conditions th

Page 49 - SAMPLE TREND OUTPUT

5-2 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual5Alarm limitsCAUTION: TO ENSURE THAT ALARM LIMITS ARE APPROPRIATE FOR THE PATIENT BEING M

Page 50

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 5-35ALARM SILENCEAudible alarms may be silenced, while visual alarms remain active. The alarm

Page 51

5-4 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual5messagesMESSAGESThe Rad-8 will indicate other data or system errors. Message conditions a

Page 52

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 5-55MESSAGES (CONTINUED)DISPLAY TYPE SOLUTIONNo Sensor ConnectedConnect sensor to cable.Senso

Page 54

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 6-16troubleshootingTroubleshootingThe following chart describes what to do if the Rad-8 syste

Page 55 - MESSAGES

6-2 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual6Troubleshooting continued PROBLEM POSSIBLE CAUSE(S) RECOMMENDATIONSpO2 VALUES DO NOT CORR

Page 56 - MESSAGES (CONTINUED)

Rad-8 Signal Extraction Pulse Oximeter Operator’s ManualiiiSAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES If using Rad-8 during full body radiati

Page 57

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 7-17specificationsRad-8 specificationsPERFORMANCEMeasurement RangeArterial Oxygen Saturation

Page 58 - Troubleshooting

7-2 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual7specificationsRad-8 specifications continuedPHYSICAL CHARACTERISTICSDimensions:8.2” x 6.

Page 59 - Troubleshooting continued

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 7-37Rad-8 specifications continuedspecifications3 Masimo SET technology with LNOP Adt sensor

Page 60 - ENVIRONMENTAL

7-4 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual7specificationsSERIAL INTERFACE SETUP (CONTINUED)The pin-outs for the RS-232 connector are

Page 61 - PHYSICAL CHARACTERISTICS

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 8-18sensors & patient cablesIntroductionThis section covers the use and cleaning of Masim

Page 62

8-2 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual8sensors & patient cablesMasimo SpO2 SensorsBefore use of any sensor or cable, careful

Page 63

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 8-38LNOP® REUSABLE SENSORS LNOP reusable sensors must be used in conjunction with Red PC cabl

Page 64 - Introduction

8-4 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual8sensors & patient cablesLNCS® REUSABLE SENSORS LNCS reusable sensors must be used in

Page 65

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 8-58SENSOR ACCURACYRefer to Section 7, Specifications for SpO2 and pulse rate accuracy, unles

Page 67

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manualivtable of contentsSECTION 1 - OVERVIEWAbout This Manual ...

Page 68

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 9-19service / maintenanceIntroductionThis chapter covers how to test the operation, properly

Page 69

9-2 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual9Battery ServiceWARNING: THE BATTERY SHOULD BE INSTALLED AND/ OR REMOVED FROM THE RAD-8 BY

Page 70

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 9-39service / maintenancePerformance verification continued LED BRIGHTNESS:1. With the monit

Page 71 - 8 begins normal operation

9-4 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual9Service and repairREPAIR POLICYMasimo or an authorized Service Department must perform wa

Page 72

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 9-59Sales & End-User License AgreementTHIS DOCUMENT IS A LEGAL AGREEMENT BETWEEN YOU (“PU

Page 73

9-6 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual9service / maintenanceExclusions continuedTHIS WARRANTY, TOGETHER WITH ANY OTHER EXPRESS W

Page 74

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 9-79service / maintenanceRestrictions1. Copyright Restrictions: The Software and the accompa

Page 76

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 10-110AccessoriesAccessoriesPART NUMBER DESCRIPTION15111 Rad-8, Monitor Bag33134 Rad-8 Operat

Page 77

www.masimo.comInstruments and sensors containing Masimo SET technology are identified with the Masimo SET logo.33116/5303B-0209

Page 78 - Accessories

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manualtable of contents vSECTION 4 - OPERATION (CONTINUED)Setup Menu Level 1 ...

Page 79 - 33116/5303B-0209

Rad-8 Signal Extraction Pulse Oximeter Operator’s ManualSECTION 6 - TROUBLESHOOTINGTroubleshooting ..............