Radical Signal Extraction Pulse Oximeter Service Manual
UM-10026 – 1204
5-2
5.3. Troubleshooting
5.3.1. The troubleshooting procedures in this manual instruct the
technician to isolate the failure down to the sub-assembly level.
The intention is to have the Unit Under Test (UUT) repaired with
minimal steps. The Masimo Set Tester is needed to perform the
Performance Verification in Section 4 upon completion of repairs.
5.3.2. Tools required are as follows:
Phillips head screwdriver #0
Multimeter
5.3.3. The troubleshooting steps below are presented in order of the
most probable cause first. Proceed to the next step only if unit
continues to fail.
5.3.4. Always note the software versions of boards prior to repair.
5.3.5. All units must have Performance Verification Section 4 performed after
repairs have been affected.
5.4. Service Messages:
Upon detection of a system error, the main software alerts the
user with an audible alarm indicator and displays the appropriate
error code on the Handheld’s main display. The MS-X board is
held in reset and no other runtime functionality is available to the
user until the Handheld is power cycled by the user (with power
on and off key) or until power is completely removed and re-
applied to the system (removal and replacement of batteries).
System Error System Error Code
MS-X Board Failure 1
MS-X Board Communication Failure 2
5V Voltage Supply Out of Range 3
Red Voltage Supply Out of Range 4
Blue Voltage Supply Out of Range 5
Battery Drop Voltage Out of Range 6
Battery Voltage Out of Range 7